Indoor Background Ionizing Radiation Levels in Three Selected Departments of Federal Teaching Hospital, Gombe, Gombe State
Keywords:
Rados survey meter, sampled location, the dose limit, cumulative meanAbstract
This study has assessed the level of indoor background ionizing radiation in the pharmacy, radiotherapy/oncology, and radiology departments of Federal Teaching Hospital, Gombe. To achieve the required objectives, the RADOS 200 survey meter was used to captured the radiation level. The meter was held 1 meter above ground level and the readings were taken at various locations (as determined by a hand GPS) within each department. The cumulative mean background radiation exposures in the pharmacy, radiotherapy/oncology and radiology departments were estimated to be 0.0891 ± 0.0126 μSv/hr, 0.0780 ± 0.0173 μSv/hr and 0.0900 ± 0.0143 μSv/hr respectively. In conclusion, results obtained in this study show that the indoor background ionizing radiation for the sampled location comply with the 0.133 μSv/hr maximum dose limit for the public as recommended by the International Commission on Radiological Protection (ICRP).